I、Summary
To make the TDS conductivity test easy ,we designed a lot of tds sensor。As the most easy test item, TDS sensor was used in homehold water purifier,commercial water purifier, fish tank,agricuture ,etc. TDS-X10-2 series sensors includes TDS210 for 2/8 inches tube and TDS310-2 for 3/8 inches tube.
II、Important Param.
notes:
1 for less than 2000us/cm3
III、Demension
DS-310-2 stucture in mm

IV、Reference design
Calculate Steps:
1 get the conductivity of liquid on the current temperature
Conductivity K=Q/R, we get the resistance of the liquid, Q is a constant of the probe.
2 Convert the conductivity to 25℃
K25=Ktest/(1+(Ttest-25)×2%)
3 get the TDS
TDS=K25 x0.5
V、Install instructions
II、Important Param.
Introduction | Min | Typical | Max | Unit | |
Electronic param. | |||||
Operating Voltage | 3.3 | 5 | 5.5 | v | |
Operating Current | 2 | 4 | mA | ||
resistance when open | when circuit is open, the res. between the two probes | 5 | 50 | 100 | mΩ |
resistance when short circuit | when circuit is shorted, the res. between the two probes | 2 | Ω | ||
Operating param. | |||||
measure range | the range when it can works | 0 | 2000 | uS/cm2 | |
0 | 1000 | ppm① | |||
TDS resolution | 0.1 | 1 | ppm | ||
Temp. calibration range | the range when temprature calibration work normal | 0 | 25 | 50 | ℃ |
probe life | based on the working condition | 1 | 2 | 3 | Year |
pressure range | max pressure for the probe | 4 | 9 | Bar | |
water velocity limit | max water velocity | 0 | 0.5 | 2 | L/min |
Storage Temp. | when the package is instant | 0 | 50 | ℃ | |
Storage Humudity | when the package is instant | 0 | 95 | %RH | |
structure param. | |||||
probe length | metal parts ouside | 5.9 | 6 | 6.1 | mm |
probe diameter | 1 | mm | |||
electrode gap | 2 | mm | |||
probe head length | TDS210 | 59 | mm | ||
TDS310 | 69 | mm | |||
wire length | standard 500mm, can be customized | 480 | 500 | 520 | mm |
notes:
1 for less than 2000us/cm3
III、Demension
DS-310-2 stucture in mm

IV、Reference design
Calculate Steps:
1 get the conductivity of liquid on the current temperature
Conductivity K=Q/R, we get the resistance of the liquid, Q is a constant of the probe.
2 Convert the conductivity to 25℃
K25=Ktest/(1+(Ttest-25)×2%)
3 get the TDS
TDS=K25 x0.5